Bayer HealthCare Rapidlab 800 User Manual

Page 622

Advertising
background image

IĆ2

800 Series Operator's Manual

Figure IĆ1.āăPotentiometric Cell

Voltmeter

Sample Solution

Reference Electrode

Measuring Electrode

Liquid Junction

Each electrode, which acts as a half-cell with a half-cell potential, contains an inner
reference element immersed in an internal electrolyte solution. The measuring
electrode is designed to respond to changes in the concentration of the specific
analyte being measured in the sample solution. It develops a half-cell potential that
is directly related to the concentration or activity of the specific analyte. The
reference electrode provides a steady, unchanging potential to the cell. Both
electrodes are connected to the measuring device. With the current in the cell at
zero, the potential developed by the electrochemical cell is determined by
calculating the difference in potential between the measuring electrode and the
reference electrode.

E

cell

= E

meas

– (E

ref

+ E

lj

)

where

E

cell

=

electrochemical cell potential

E

meas

=

measuring electrode half-cell potential

E

ref

=

reference electrode half-cell potential

E

lj

=

liquid junction potential

The liquid junction potential (E

lj

), a small but significant voltage, develops at the

liquid junction between the reference electrode, which contains a solution of
saturated potassium chloride, and the sample solution. This potential occurs
because of the different rates at which chemical species diffuse across the
boundary between two liquids. This difference in rates results in a charge
separation that gives rise to the liquid junction potential. Although the potential
formed is small, it must be considered when measuring cell potential.

9