Checking the solution levels in the sensors – Bayer HealthCare Rapidlab 800 User Manual

Page 185

Advertising
background image

3Ć17

Maintaining the System

Checking the Solution Levels in the Sensors

Use this procedure to ensure that the sensors contain fill solution to the levels
described in Figure 3-5.

Figure 3Ć5ā.ăFill Solution Levels

Sensor

Fill Solution Level

pH, K

+

, Cl

, and Ca

++

nearly full and with a bubble at the top

Na

+

to the top

Reference

to the fill line

NOTE:

The pO

2

and pCO

2

sensors do not require you to maintain them, even

though they have fill solution. Slight discoloration of the fill solution in the pO

2

and pCO

2

sensors is normal.

NOTE:

The glucose and lactate biosensors do not require fill solution.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Open the measurement module door as shown in Figure 3-6:

a. Push up the latches on the measurement module door.

b. Lift the door.