Troubleshooting the System

c. Attach another piece of test tubing to the waste tubing connector on the

CO-ox pump tubing, and place the test tubing in the graduated cylinder.

6.

Select the pump and speed to test and press

Enter

.

7.

Prime the tubing:

a. Press

Start Test

.

b. Allow reagent water to flow through all the tubing.

c. Press

Stop Test

.

d. Empty the graduated cylinder.

8.

Test the pump flow rate for 1 minute:

a. Place the outlet tubing in the graduated cylinder.

b. Press

Start Test

and simultaneously start timing with a stopwatch.

c. Collect fluid for exactly 1 minute.

d. Press

Stop Test

.

e. Compare the volume of collected fluid to the expected ranges.

Pump

Volume at low speed

Volume at high speed

reagent

1.51 – 1.85 mL/min

5.94 – 7.26 mL/min

sample

1.51 – 1.85 mL/min

5.94 – 7.26 mL/min

waste

1.51 – 1.85 mL/min

7.02 – 8.58 mL/min

CO-ox

0.27 – 0.33 mL/min

5.94 – 7.26 mL/min

9.

If the measured volume is not acceptable, the pump tubing may be worn or
damaged, the roller cage or shaft may be dirty, or the pump speed may require
calibration.

10. Perform the appropriate action.

If you want to . . .

Then . . .

measure the flow rate of the same
pump at a different speed

a. Empty the graduated cylinder.

b. Select the pump and speed to test and press

Enter

.

c. Repeat steps 7 and 8.

measure the flow rate of a
different pump or return to the
Menu screen

continue with step 11.