D3 slope error – Bayer HealthCare Rapidlab 800 User Manual

Page 304

Advertising
background image

4Ć20

800 Series Operator's Manual

D3 Slope Error:

tHb

Problem

The tHb slope is beyond predefined limits during the calibration.

Qualifiers:

tHb

Solutions

Ensure the target value entered for the tHb slope is correct for the reagent used.
If required, re-enter the value and repeat the calibration.

Ensure the slope reagent is appropriate for the instrument.

Check that the sample chamber is tightly in place. The tab on the cams should
be vertical. Repeat the tHb slope calibration.

Check that the gasket in the sample chamber is seated correctly. Refer to
Cleaning the Sample Chamber, in Section 3. Replace the gasket, if necessary.

If the D code reappears, contact your Service Representative.

D3