EĆ14

800 Series Operator's Manual

Table EĆ12.āă840 Recovery and Precision TestingĊpCO

2

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

24

0.08

14.0

14.3

98.0

Capillary

18

0.32

13.6

14.3

95.1

Microcapillary

24

0.34

14.0

14.3

97.6

Microsyringe

24

0.13

14.3

14.3

99.9

Expired Gas

24

0.10

14.5

14.3

101.4

2

Syringe

24

0.14

21.6

21.4

100.8

Capillary

18

0.17

21.4

21.4

100.0

Microcapillary

24

0.45

21.4

21.4

100.2

Microsyringe

24

0.15

21.6

21.4

101.2

Expired Gas

24

0.14

21.6

21.4

100.7

3

Syringe

24

0.22

35.8

35.7

100.4

Capillary

18

0.53

35.3

35.7

98.8

Microcapillary

23

0.77

36.2

35.7

101.3

Microsyringe

24

0.30

35.7

35.7

99.9

Expired Gas

24

0.25

35.5

35.7

99.4

4

Syringe

24

0.28

50.1

49.9

100.3

Capillary

18

0.40

51.3

49.9

102.8

Microcapillary

24

1.28

51.5

49.9

103.2

Microsyringe

24

0.21

50.3

49.9

100.8

Expired Gas

24

0.37

49.4

49.9

99.0

5

Syringe

24

0.26

71.5

71.3

100.3

(Continued)