System Administration

5Ć15

If you change parameter units for pH after the system has started to collect data,
QC statistics are still computed in the current units.

Defining Parameter Values

Use this procedure to define the values for ctHb and p50 that the system uses when
no value is entered or measured, and to define the value for the O

2

binding factor

and ctO

2

(a-v), which are constants supplied by the system. The system uses these

values for certain calculations. Table 5-5 lists the default values and valid ranges
for ctHb, O

2

binding factor, p50, and ctO

2

(a-v).

Table 5Ć5.āăDefault Measurement Values

Parameter

Valid Range

Default Value

Units

ctHb

2.0 – 25.0

20 – 250

1.2 – 15.5

15.0

g/dL

g/L

mmol/L

p50

0.0 – 100.0

0.00 – 13.32

26.5

mmHg

kPa

O

2

Binding

1.35 – 1.39

1.39

ctO

2

(a-v)

0.0 – 20.0

0 – 200

0.0 – 9.0

3.5

mL/dL

mL/L

mmol/L

1.

Access the Units and Values screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

4 Units/Values

and press

Enter

.

The first Units and Values screen appears.

2.

Press

Next Screen

.

The second Units and Values screen appears as shown in Figure 5-7.

Procedural

Notes

5

Menu Code

4