Installation

HĆ13

48. Verify sensor performance by completing two successful two-point

calibrations.

The Ready screen appears when the calibration finishes.

49. Analyze QC material and verify acceptable results.

50. Replace the front cover.

Attaching the 800 COĆoximeter Module to the 800 Base

Model

BIOHAZARD:

Refer to Appendix A for recommended precautions when working

with biohazardous materials.

The 800 series CO-oximeter (CO-ox) module must be installed by an authorized
Bayer Diagnostics Representative. Unauthorized use of these procedures can void
the warranty or service contract.

For detailed information about the software, refer to Learning About the System in
Section 1 and Operating the System in Section 2 of the operator’s manual.

Select a location for the new system configuration where it is not exposed to direct
sunlight. Refer to Table H-3 for system specifications.

Table HĆ3.āăSystem Specifications

Property

Specification

ambient operating
temperature

15 – 32

°

C

ambient operating relative
humidity

5 – 85%, non-condensing

power rating

400VA (maximum)

power requirements from
the 800 base model:

100V (

±

) 10% 50/60 Hz

120V –15 to +10% 50/60 Hz

220V (

±

) 10% 50/60 Hz

240V (

±

) 10% 50/60 Hz

(Continued)