Bayer HealthCare Rapidlab 800 User Manual

Page 254

Advertising
background image

3Ć82

Rapidlab 800 Operator's Manual

CAUTION:

Do not touch the internal electrode wire. The wire is fragile and is

easily damaged.

3.

Unscrew the internal electrode and carefully set it aside on a lint-free tissue as
shown in Figure 3-57.

Figure 3Ć57ā.ăRemoving the Internal Electrode

Internal Electrode

OĆring

Electrode

Compartment

OĆring

4.

Empty the fill solution remaining in the sensor.

NOTE:

Refer to Table 3-1 to determine the appropriate fill solution to use.

5.

Rinse the electrode compartment with 3 drops of the appropriate sensor fill
solution and then empty the fill solution from the sensor.

6.

Slowly add the appropriate fill solution as shown in Figure 3-58:
S Fill the pH, K

+

, Cl

, and Ca

++

sensors almost full, leaving a bubble at the

top.

S Fill the Na

+

sensor to the top.

!