Operating the System

Figure 2Ć23.ā Analyzing a QC Sample With an Aspiration Adapter

Sample Port

Aspiration Adapter

QC Sample

CAUTION:ĄĂ

To ensure the accuracy of the CO-ox measurement, close the

CO-ox cover before pressing Analyze.

4.

Press

Analyze

.

5.

When prompted, remove the sample device.

The QC File Information screen appears.

If . . .

Then . . .

you already
scanned the
ampule

the QC File Information screen is complete. Continue with
step 7.

Auto ID is on

the system completes the data entry form automatically at the
end of analysis. The data entry form remains on the screen
until you press Done.

Press

Done

to assign the results to the file. The system then

displays the final results and prints the report. Continue with
step 8.

you did not scan
the ampule and
Auto ID is off

the QC File Information screen remains displayed until you
complete all required fields and press Done. Continue with
step 6.

6.

Type the file number or scan the bar code from the ampule, and press

Done

.

When you enter the file number, the system scans the QC database for that file
number. If a match is found, the system completes the remaining fields except
the Operator ID field.

Refer to Entering QC Sample Data, page 2-43, if you need more information
to complete the QC File Information screen.

!