3Ć77

Maintaining the System

Replacing the Internal Reference Electrode

Materials required:

S reference electrode internal replacement kit
S hex tool
S lint-free tissue

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

CAUTION:ĄĂ

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

2.

Remove the reference sensor as shown in Figure 3-51:

a. Push up the latches on the measurement module door and lift the door.

b. Push the spring-loaded latch to the right.

c. Grasp the tab on the reference sensor and pull the sensor up and out of the

measurement module.

Figure 3Ć51.āăRemoving the Reference Sensor from an 850

Grasp the tab on the reference

sensor and pull it up and out of the

measurement module.

Push the springĆloaded latch

to the right.

3.

Use the hex tool to remove the internal electrode from the reference sensor as
shown in Figure 3-52.

2

Menu Code

7

!