Rapidlab 800 Operator's Manual

Figure 3Ć21.āăReplacing the Measurement Module Tubing

Position 3 Connector

Tubing Groove

Measurement Module

Tubing

Connector at the

SpringĆLoaded Latch

7.

Close the measurement module door.

8.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

9.

Press

Yes

to perform a two-point calibration.