Maintaining the System

5.

Replace the measurement module tubing:

a. Remove the tubing from the connector at the spring-loaded latch in the

measurement block as shown in Figure 3-20.

Figure 3Ć20.āăReplacing the Measurement Module Tubing on an 840

Position 3 Connector

Tubing Groove

Measurement Module Tubing

Connector at the

SpringĆLoaded Latch

Spacer

b. Remove the tubing from the spacer.

c. Disconnect the tubing from the connector at position 3 on the reagent

manifold.

d. Push the new tubing through the spacer.

e. Connect the left end of the tubing to the spring-loaded latch.

f. Connect the right end of the tubing to the connector at position 3.

g. Press the tubing into the tubing groove.

h. Continue with step 7.

6.

Replace the measurement module tubing:

a. Remove the tubing from the connector at the spring-loaded latch in the

measurement block as shown Figure 3-21.

b. Disconnect the tubing from the connector at position 3 on the reagent

manifold.

c. Connect the left end of the new tubing to the spring-loaded latch.

d. Connect the right end of the tubing to the connector at position 3.

e. Press the tubing into the tubing groove.