Ph sensor, Carbon dioxide tension (pco – Bayer HealthCare Rapidlab 800 User Manual

Page 629

Advertising
background image

Operating Principles

IĆ9

pH is clinically significant as a means of determining acid-base disturbances.
Acid-base disorders can result in several pathologic conditions. An acid-base
disorder resulting initially from ventilatory dysfunction is called a primary
respiratory acidosis or alkalosis, while a disorder due to renal or gastrointestinal
inadequacy is referred to as metabolic acidosis or alkalosis. Using acceptable
therapeutic ranges, a pH less than 7.3 indicates acidosis, and a pH greater than 7.5
indicates alkalosis.

14

pH Sensor

The pH sensor, which is based on ISE technology, is a half-cell that forms a
complete electrochemical cell when combined with the external reference sensor. It
contains a silver/silver chloride wire surrounded by a buffer solution. A glass
membrane that is highly sensitive and specific for hydrogen ions separates the
sample from the solution.

Figure IĆ5.āăpH Sensor

Silver/Silver Chloride
Wire

Buffer Solution

Sample Path

As the sample comes in contact with the membrane of the pH sensor, a membrane
potential develops due to the exchange of hydrogen ions in the membrane. The
silver/silver chloride inner conductor transmits the potential to a voltmeter where it
is compared to the constant potential of the reference sensor. The final measured
potential reflects the hydrogen ion concentration of the sample and is used to
report the pH value of the sample.

Carbon Dioxide Tension (pCO

2

)

Carbon dioxide (CO

2)

is produced during normal cell metabolism and is released

into the blood stream where it is transported to the kidneys and lungs for excretion.
CO

2

is transported through the blood as bicarbonate (HCO

3

), dissolved CO

2

, and

carbonic acid (H

2

CO

3

). CO

2

exists in a dynamic state in the blood as seen in the

following equation:

CO

2

+ H

2

O

{ H

2

CO

3

{ HCO

3

+ H

+

Advertising