2Ć17

Operating the System

7.

When prompted, remove the sample device.

8.

Type the required information in the Patient Information screen and then press

Done

.

Refer to Entering Patient Sample Data, page 2-27, if you need more
information to complete this screen.

Results appear on the screen and are continuously updated until analysis is
complete. The system then displays the final results and prints the report.

9.

The system performs a wash at the end of analysis and then returns to the
Ready screen. If the system continues to display the Results screen after the
wash finishes, press

Home

to return to the Ready screen.

If the system analyzes an insufficient sample, the message, Insufficient Sample,
prints on the roll printer report.

The system performs a wash and discards any data for the sample if any of the
following occur:

S You press Cancel

at any time to interrupt a sample analysis.

S You do not press Sample in Place within 5 minutes of the screen prompt.
S You do not remove the capillary tube within 5 minutes of the screen prompt.

Procedural

Notes