Replacing the printer paper – Bayer HealthCare Rapidlab 800 User Manual

Page 238

Advertising
background image

3Ć66

Rapidlab 800 Operator's Manual

9.

Perform a two-point calibration.

a. Press

Calibrate

.

b. Select Two-point and press

Enter

.

c. Press

Start Calibration

.

Perform two-point calibrations after changing the calibration reagents to ensure
that the reagents are acceptable and the system is functioning properly. You can
analyze quality control materials with the new reagents and compare the results
with the previous QC results after the system is recalibrated.

Replacing the Printer Paper

CAUTION:

Do not attempt to print without paper installed or with a double

thickness of paper in the printer. Use the paper advance key to feed the paper
through the printer.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Pull up the paper spool.

3.

Firmly grasp the old roll of paper and pull it off the spool.

NOTE:

Grasp the roll of paper tightly to slide the paper completely off the spool.

4.

Reinstall the paper spool.

5.

Lift the printer cover.

6.

Push up the printer lever.

7.

Install a roll of printer paper:

a. Unroll a small amount of paper from the new roll and cut or fold a clean

straight edge, if necessary.

b. Place the paper roll in the cavity with the paper unrolling from the bottom

as shown in Figure 3-38.

Procedural

Notes

!

2

Menu Code

7