4Ć60

800 Series Operator's Manual

c. Attach another piece of test tubing to a two-way connector, attach the

connector to the outlet tubing for the appropriate pump, and place the test
tubing in the graduated cylinder.

If you want to . . .

Then connect the test tubing . . .

test the reagent pump

a. to the pump inlet tubing connected to right position 1

b. and to the outlet tubing connected to left position 1.

test the sample pump

a. to the pump inlet tubing connected to left position 4

b. and to the outlet tubing connected to right position 4.

test the waste pump

a. to the pump inlet tubing connected to left position 5

b. and to the outlet tubing connected to right position 5.

Figure 4Ć5.ā Measuring the Flow Rate for the Sample or Waste Pump

Sample or
Waste Pump

Two-Way
Connector

Tubing

Tubing

Two-Way
Connector

Reagent
Water

5.

Install the test tubing on the CO-ox pump:

NOTE:ĄĂ

The CO-ox pump turns clockwise.

a. Disconnect the sample and waste tubing from the pump tubing.

b. Attach one piece of test tubing to sample tubing connector on the CO-ox

pump tubing, and place the test tubing in a beaker of reagent water.