Cleaning the roll printer – Bayer HealthCare Rapidlab 800 User Manual

Page 226

Advertising
background image

3Ć56

Rapidlab 800 Operator's Manual

NOTE:

Ensure that the three O-rings are in place.

7.

Reinstall the sample port, matching the tab on the sample port to the notch in
the retainer ring.

8.

Push the tab on the retainer ring away from you until it locks in place.

9.

Reinstall the sample door, ensuring that it snaps in place.

10. Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

11. Press

Yes

to perform a two-point calibration.

Cleaning the Roll Printer

Materials required:

S alcohol
S standard copier paper

WARNING

Wear gloves to protect your hands from the alcohol.

1.

Remove the paper roll from the roll printer:

a. Pull the paper spool up and remove the paper from the spool.

b. Lift the printer lever and remove the paper from the printer.

c. Remove the paper roll and set it aside.

2.

Cut a piece of standard copier paper in half lengthwise to make a strip of paper
10.8 cm (4.25 inches) wide.

3.

Moisten the center of the paper along its width with the alcohol. Leave the two
ends of the paper dry.

4.

Shut down the system from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

3 Shutdown

and press

Enter

.

c. Press

Yes

.

CAUTION:ĄĂ

You must wait at least 1 minute before you disconnect the power

cord, and then wait at least 10 seconds before you reconnect the power cord. If
you do not adhere to the time intervals, you can damage the system.

5.

Wait at least 1 minute, and then disconnect the power cord from the power
supply.

6.

Insert the paper into the printer, and push the paper through the printer until
the moistened part of the paper is past the print head.

7.

Pull the paper back and forth several times past the printer head.

7

Menu Code

3

!

Advertising