Bayer HealthCare Rapidlab 800 User Manual

Page 220

Advertising
background image

3Ć50

Rapidlab 800 Operator's Manual

6.

Ensure that the O-ring is in place. Replace the O-ring if it is worn or damaged.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

NOTE:

If the sensor does not insert easily into the measurement module, slide the

remaining sensors to the right to create more space.

7.

Reinstall the sensor:

a. Gently align the top of the sensor with the sensor contact.

b. Snap the body of the sensor down into place.

c. Press the tab on the spring-loaded latch to release the latch.

8.

Close the measurement module door.

9.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

10. Press

No

.

11. Allow the system to warm up for at least 15 minutes.

12. Check the sample path temperature from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

2 Temp/pAtm

and press

Enter

.

c. Press

Start Test

.

d. Press

Stop Test

.

e. Check the screen for the sample path temperature.

f. If the temperature control system is off, press

Reset Control

.

g. Press

Exit Test

.

13. Perform a two-point calibration from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

!

3

Menu Code

2

(from the Main Menu)

1

Menu Code

2

(from the Main Menu)