Bayer HealthCare Rapidlab 800 User Manual

Page 195

Advertising
background image

3Ć27

Maintaining the System

Figure 3Ć12.āăRemoving the Sample Chamber

Sample Chamber

Sapphire
Window

Mounting
Pin

Cam

Bubble Trap

4.

Remove the gasket and support:

a. Slide your finger under the flat metal support of the gasket.

b. Lift the gasket and support from the sample chamber.

Figure 3Ć13.āăRemoving the Gasket

Gasket Support

Gasket

Mounting Pin
Holes

Inlet

Outlet

CAUTION:

Do not touch the optical zone (front or back area enclosed by the

gasket) of the sample chamber with your fingers.

Figure 3Ć14.āăOptical Zone

Optical Zone Areas

Sample Side

Port Side

5.

Push a 0.016-inch diameter clot-removal line into the metal tubing of the inlet
and outlet ports of the sample chamber.

6.

Flush the inlet and outlet ports to remove any clots or debris:

!