Cleaning the sample path with bleach – Bayer HealthCare Rapidlab 800 User Manual

Page 229

Advertising
background image

3Ć59

Maintaining the System

8.

Perform a two-point calibration from the Menu screen:

NOTE:ĄĂ

Glutaraldehyde may cause excessive drift to the K

+

and Ca

++

sensors.

Repeat calibrations until the system performs a successful calibration.

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

9.

Analyze a minimum of two levels of quality control material to verify sensor
performance.

10. Empty and clean the waste bottle, waste outlets, and waste outlet cover as

described in

Emptying the Waste Bottle

, page 3-45.

Cleaning the Sample Path with Bleach

Materials required:

S 10% solution of household bleach

NOTE:

Dilute household bleach (5.25% sodium hypochlorite) 1:10 with

reagent-quality water.

S aspiration adapter
S lint-free tissue or swabs
S test/blank reference sensor (TB5)
S test/blank glucose sensor (TB4)
S test/blank lactate sensor (TB4)