And p co, Are measured, 1ć69 learning about the system – Bayer HealthCare Rapidlab 800 User Manual

Page 91: Continued)

Advertising
background image

1Ć69

Learning About the System

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

7.3

6.8

Wash Clean

5

CO

H

SC

3.

After the wash segment moves through the measurement
module, the Cal gas and diverter valves open and Cal gas
flows into the inner sample port. The sample pump
aspirates the Cal gas at a controlled rate through the
measurement module where pO

2

and pCO

2

are measured.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

7.3

6.8

Wash Clean

5

CO

H

SC

4.

The Slope gas valve opens and Slope gas flows into the
inner sample port. The sample pump aspirates the Slope
gas at a controlled rate through the measurement module
where pO

2

and pCO

2

are measured. The CO-ox pumps

starts to prevent contamination of the sample connector by
ambient air.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

• • • • • • • • • • •

••

• • • • •

7.3

6.8

Wash Clean

5

CO

H

SC

5.

After the gases are measured, the 7.3 valve, wash bypass
valve, and diverter valve open. The reagent pump moves a
7.3 reagent pre-segment and main segment from the
reagent manifold toward the inner sample port. The
sample pump is on.

(Continued)