Replacing a roller cage – Bayer HealthCare Rapidlab 800 User Manual

Page 268

Advertising
background image

3Ć96

Rapidlab 800 Operator's Manual

Replacing a Roller Cage

Use this procedure to replace the roller cage for the reagent, sample, waste, or
CO-ox pump.

Materials required:

S reagent water
S lint-free tissue
S roller cage

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Disconnect the pump tubing from the reagent manifold.

NOTE:

Do not stretch the tubing.

3.

While holding the left side of the tubing, turn the roller cage clockwise and
gently pull the tubing away from the platen and the roller cage.

4.

Set the tubing assembly aside.

5.

Grasp the roller cage and gently pull it straight off its shaft as shown in
Figure 3-65.

2

Menu Code

7

Advertising