Bayer HealthCare Rapidlab 800 User Manual

Page 469

Advertising
background image

System Administration

5Ć59

Table 5Ć23.āăCalibration Gas Values

Gas

Cal Range

Cal Default

Slope Range

Slope Default

pCO

2

2.00 – 7.90%

5.00%

(Cal CO

2

+ 2.9) –

99.9%

10.00%

pO

2

5.00 – 98.00%

12.00%

0.00% (fixed)

0.00%

1.

Access the Calibration Gas Values screen from the Menu screen:

a. Select

5

Operating Setup

and press

Enter

.

b. Select

6

Calibration Setup

and press

Enter

.

c. Select

3

Gas Values

and press

Enter

.

The Calibration Gas Values screen appears, as shown in Figure 5-36.

Figure 5Ć36.ā Calibration Gas Values Screen

You cannot change

pO

2

slope.

Press Clear Entry to

delete text in a field.

2.

Move the cursor to the field that you want to change.

3.

Type the new gas value and press

Enter

.

4.

Repeat steps 2 and 3 for each gas value you want to change.

5.

Press

Done

.

6.

Define another setup function or press

Exit Menu

to return to the Ready

screen.