Replacing the coćox lamp – Bayer HealthCare Rapidlab 800 User Manual

Page 275

Advertising
background image

3Ć103

Maintaining the System

8.

Pinch the ends of the clamp and pull it away from the bracket.

9.

Remove the old lamp from the clamp and discard it.

CAUTION:

Avoid touching the lamp with your fingers.

10. Place a new lamp in the clamp with the front of the lamp facing the system.

11. Pinch the ends of the clamp together and reinstall it in the bracket.

12. Reinstall the connector in the system.

13. Reinstall the lamp cover by inserting the tabs on the cover into the system and

snapping the cover back into place.

14. Reconnect the power cord to the power source and allow the system to warm

up for at least 15 minutes.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

15. Press

No

.

16. Verify the system component temperatures from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

2 Temp/pAtm

and press

Enter

.

c. Press

Start Test

.

d. Verify that the system components warm up.

e. Press

Stop Test

.

f. If the temperature control system is off, press

Reset Control

and allow the

system to warm up.

g. Press

Exit Test

.

17. When the temperature is stable, perform a two-point calibration from the

Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

Replacing the COĆox Lamp

WARNING

Burn Hazard. Do not touch or look at the lamp when the lamp is

illuminated. Injury to the skin or the eye can occur when the lamp is operated
outside the lamp housing.

Materials required:

S CO-ox lamp
1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

!

3

Menu Code

2

1

Menu Code

2

(from the Main Menu)

2

Menu Code

7