Bayer HealthCare Rapidlab 800 User Manual

Page 276

Advertising
background image

3Ć104

Rapidlab 800 Operator's Manual

b. Select

7 Stop System

and press

Enter

.

2.

Allow the lamp to cool for at least 5 minutes.

WARNING

Ensure that the lamp has been off for at least 5 minutes to allow

sufficient time for it to cool.

3.

Remove the lamp cover from the CO-ox module.

4.

Remove the lamp connector, as shown in Figure 3-71.

Figure 3Ć71ā.ăReplacing the COĆox Lamp

Lamp Cover

Lamp

COĆox Pump

5.

Remove the old lamp from the lamp housing and discard it.

6.

Place a new lamp in the housing aligning the lamp with the guide pin.

7.

Pinch the ends of the clamp together and replace it in the bracket.

8.

Install the lamp connector.

9.

Reinstall the lamp cover.

10. Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

11. Press

No

.

12. Calibrate the new lamp:

a. Select

3 Troubleshooting

and press

Enter

.

Menu Code

3

3

2