Bayer HealthCare Rapidlab 800 User Manual

Page 548

Advertising
background image

EĆ22

800 Series Operator's Manual

Level

Recovery

Expected

Observed

WRSD

*

n

Mode

Capillary

24

1.20

71.7

71.3

100.6

Microcapillary

18

1.36

71.1

71.3

99.7

Microsyringe

18

0.55

70.5

71.3

98.9

Expired Gas

18

0.49

71.0

71.3

99.6

* WRSD = within-run standard deviation

Table EĆ18.āă850 Recovery and Precision TestingĊpO

2

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

21

0.16

28.2

28.5

98.9

Capillary

21

0.18

28.9

28.5

101.4

Microcapillary

18

0.35

29.0

28.5

101.8

Microsyringe

18

0.25

28.5

28.5

100.2

Expired Gas

18

0.16

27.9

28.5

98.0

2

Syringe

24

0.22

50.5

49.9

101.2

Capillary

18

0.34

49.2

49.9

98.6

Microcapillary

24

0.41

49.4

49.9

99.0

Microsyringe

24

0.13

49.8

49.9

99.8

Expired Gas

24

0.10

49.4

49.9

99.1

3

Syringe

18

0.34

85.3

85.6

99.6

Capillary

22

0.33

85.5

85.6

99.9

Microcapillary

18

0.99

84.5

85.6

98.7

Microsyringe

18

0.70

85.6

85.6

100.0

Expired Gas

18

0.20

84.7

85.6

98.9

(Continued)