D5 No Endpoint:

Glu

Lac

Problem

Sensor does not reach stable reading during predefined time limit.

Qualifiers:

Glu

Lac

Solutions

NOTE:ĄĂ

If D codes also exist for reagent problems such as D23, D24, D29, or

D50, correct these problems first, then correct the D5.

Perform a metabolite two-point calibration.

Perform the Measurement Test for the affected biosensor as described in
Measurement Test, page 4-70.

If the glucose or lactate biosensors have been on the system less than one day,
then perform three two-point calibrations.

Replace the biosensor as described in Replacing the Glucose and Lactate
Biosensors
in Section 3.

If the D code reappears, contact your Service Representative.