Selecting automatic qc file assignment – Bayer HealthCare Rapidlab 800 User Manual

Page 463

Advertising
background image

System Administration

5Ć53

Selecting Automatic QC File Assignment

Use this procedure to control the way QC results are assigned to a QC file.
Table 5-19 describes the options for assigning QC results.

Table 5Ć19.āăAuto Identify Options

Option

Description

On

The system automatically determines the appropriate QC file
to assign QC sample results, by comparing the QC sample
results for pH and pCO

2

to the ranges entered in setup for

these parameters.

NOTE:

If the pH, pCO

2

,

or tHb sensor is disabled or not in

calibration, Auto ID does not work.

Off

The operator assigns the QC sample results to the appropriate
file. The default value is Off.

1.

Access the Global QC Settings screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

1 QC Setup

and press

Enter

.

c. Select

2 Global QC Settings

and press

Enter

.

The Global QC Settings screen appears with the cursor in the Auto ID frame,
as shown in Figure 5-31.

Figure 5Ć31.ā Global QC Settings Screen

Select On or Off from

the Auto ID field and

press Enter.

2.

Select

On

or

Off

and press

Enter

.

3.

Press

Done

.

4.

Define another setup function or press

Exit Menu

to return to the Ready

screen.