Shutting Down and Restarting the System

Use this procedure to shut down the system before you perform service and to
restart the system when finished.

1.

Access the Shutdown screen from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

3 Shutdown

and press

Enter

.

2.

Press

Yes

.

A message appears on the screen and on the roll printer, prompting you to wait
before you disconnect the power.

CAUTION:ĄĂ

Do not unplug the system until the following message appears on

the screen:

...synching disks... done

This operating system message indicates that the system has successfully
completed the shutdown procedures. Unplugging the system before this
message appears can damage the system.

3.

When you see this message, disconnect the power cord from the power supply.

CAUTION:ĄĂ

Adhere to the time interval or damage can occur to the system.

4.

Wait at least 10 seconds after disconnecting the power cord, then reconnect the
power cord into the power source.

After a few moments, the system starts initializing. When initializing is
complete, a screen for Analyze Mode appears.

In the event of a power failure or a power surge and the system shut down,
perform a two-point calibration when the system completes initializing.