Bayer HealthCare Rapidlab 800 User Manual

Page 617

Advertising
background image

Installation

HĆ25

NOTE:

Do not insert swabs into the sample port or spray anything into the

measurement module.

2.

Rinse the exterior surfaces with reagent water.

3.

Clean spills around any of the roller cages, if required:

a. Remove the roller cage as described in Replacing a Roller Cage in

Section 3.

b. Clean the roller cage and the roller cage shaft with a 10% solution of

household bleach.

c. Rinse with reagent water and dry thoroughly.

d. Reinstall the roller cage.

4.

Empty and discard the waste bottle. Clean waste outlets and waste outlet cover
as described in Emptying the Waste Bottle in Section 3.

5.

Remove the printer paper. Refer to Replacing the Printer Paper in Section 3.

6.

Tape the printer cover in place.

Shutting Down the System

1.

Shut down the system from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

3 Shutdown

and press

Enter

.

c. Press

Yes

.

The Shutdown screen appears.

CAUTION:ĄĂ

Do not unplug the system until the following message appears on

the screen:

...synching disks... done

This is an operating system message, indicating that the system has
successfully completed the shutdown procedures. Unplugging the system
before this message appears can damage the system.

2.

When you see the operating system message, you can disconnect the power
cord from the power supply.

7

Menu Code

3

!