Performance Characteristics

EĆ13

Table E-11 through Table E-13 summarize the results of the 840 system whole
blood and expired gas recovery precision testing.

Table EĆ11.āă840 Recovery and Precision TestingĊpH

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

15

0.002

7.251

7.251

100.0

Capillary

12

0.003

7.255

7.258

100.0

Microcapillary

17

0.006

7.261

7.256

100.1

Microsyringe

18

0.003

7.248

7.238

100.1

pH Only

15

0.004

7.235

7.235

100.0

2

Syringe

12

0.003

7.427

7.420

100.1

Capillary

14

0.003

7.429

7.423

100.1

Microcapillary

12

0.004

7.380

7.374

100.1

Microsyringe

15

0.002

7.400

7.397

100.0

pH Only

12

0.007

7.411

7.410

100.0

3

Syringe

15

0.002

7.663

7.654

100.1

Capillary

12

0.002

7.656

7.650

100.1

Microcapillary

18

0.007

7.664

7.664

100.0

Microsyringe

18

0.003

7.651

7.645

100.1

pH Only

15

0.004

7.636

7.635

100.0

* WRSD = within-run standard deviation