Bayer HealthCare Rapidlab 800 User Manual

Page 211

Advertising
background image

3Ć43

Maintaining the System

6.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

7.

Press

No

.

8.

Calibrate the reagent pump from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

2 Pump Flow Rate

and press

Enter

.

d. Press

Calibrate Pump

.

e. Press

Exit Test

when the Pump calibration complete message appears.

9.

Perform a two-point calibration from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

2

1

Menu Code

3

(from the Main Menu)

1

Menu Code

2

(from the Main Menu)