Bayer HealthCare Rapidlab 800 User Manual

Page 613

Advertising
background image

Installation

HĆ21

S lint-free tissue and swabs
S aspiration adapter
S test/blank sodium or pH sensor (TB2)
S glucose test/blank sensor

(TB4)

S lactate test/blank sensor

(TB4)

S reference test/blank sensor (TB5)
S valve wrench

BIOHAZARD:

Refer to Appendix A for recommended precautions when working

with biohazardous materials.

Cleaning the Sample Path with Bleach

CAUTION:

Exposure to bleach damages the glucose and lactate biosensor

membranes. Replace the glucose and lactate biosensors with the test/blank (TB4)
sensor before cleaning the sample path.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

1.

Take the appropriate action.

If you have an . . .

Then . . .

840 or 850

Go to step 2.

860

a. Remove the glucose and lactate biosensors.

b. Install the test/blank sensors.

c. Go to step 2.

2.

Deproteinize the sample path:

a. Prepare the deproteinizer as directed on the package.

b. Select

2 Maintenance

and press

Enter

.

c. Select

1 Deproteinize

and press

Enter

.

d. Invert the deproteinizer several times to mix.

e. Insert an aspiration adapter into the sample port and insert the other end

into the deproteinizer.

f. Press

Analyze

.

g. Remove the adaptor when prompted.

h. Wait 5 minutes for the deproteinizing cycle to finish.

!

860

!

2

Menu Code

1