Troubleshooting the System

Removing Obstructions from the Sample Entry Components

Use this procedure to remove obstructions from the sample entry components.
Perform the steps in sequence until you locate and remove the obstructions.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

2.

Remove obstructions from the sample entry area:

a. Place a gauze pad in front of the sample port.

b. Turn the sample pump counterclockwise until the obstruction is ejected

onto the gauze pad.

3.

Remove obstructions from the sample tubing:

a. Inspect the sample tubing for obstructions.

b. Remove the sample tubing.

c. Push a 0.558 mm (0.022-inch) diameter clot-removal line through the

sample tubing from back to front, opposite the direction of sample flow, as
shown in Figure 4-13.

d. Reinstall the sample tubing.

Figure 4Ć13.ā Removing Obstructions from the Sample Tubing

Clot-removal
Line

Sample Tubing