Learning About the System

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

7.3

6.8

Wash Clean

5

CO

H

SC

9.

A small amount of foam wash is moved to the
measurement module to remove the 7.3 reagent. The
reagent pump then moves three 6.8 reagent pre-segments
and a main segment to the inner sample port. The
pre-segments are used to remove any residual 7.3 reagent
from the pathways.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

• • • • • • • • • • • • • • • • • • • • • • • •

•••••••••••••••••

••

• • • • •

7.3

6.8

Wash Clean

5

CO

H

SC

10. The pre-segments move through the sample port into the

waste.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

• • • • • • • • • • • • • • • • • • • • • • • •

• • • • • • • • • • •

••••••••••••••••••••

•••

••••••••••••••••••

• • •

• • • • •

7.3

6.8

Wash Clean

5

CO

H

SC

11. When the main segment reaches the sample port, the

sample pump starts and moves the segment to the
measurement module until it is detected by fluid detector 1
and 1A. The main segment moves toward the sample port.
The waste pump starts.

(Continued)