800 Series Operator's Manual

5Ć54

Selecting Automatic Acceptance of QC Results

Use this function if you are connected to an LIS or data management system and
you want the 800 system to automatically accept QC results. When you use Auto
Accept with Auto Send, the 800 system accepts all QC samples and sends them to
the connected LIS or data management system. You can review the QC results at
the LIS or data management system and, if required, change the status to reject or
discard.

NOTE:

Auto Send must be turned on to have results automatically sent to an LIS

or data management system.

Table 5Ć20.āăAuto Accept QC Options

Option

Description

On

The system automatically accepts QC results and, when Auto Send is on,
sends the results to an LIS or data management system.

Off

The operator accepts, rejects, or discards QC results at the end of analysis.
The default setting is Off.

1.

Access the Global QC Settings screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

1 QC Setup

and press

Enter

.

c. Select

2 Global QC Settings

and press

Enter

.

The Global QC Settings screen appears, as shown in Figure 5-32.

Figure 5Ć32.ā Global QC Settings Screen

Move to the Auto

Accept QC frame and

select On or Off.

2.

Move to the Auto Accept QC frame.

3.

Select

On

or

Off

and press

Enter

.

4.

Press

Done

.

2

1