System Administration

5Ć33

Changing the Auto Clean Time

Use this procedure to change the time when the system performs the automatic
cleaning of the reagent manifold. Auto Clean occurs once every 24 hours. The
default time is 0200.

On the first day of the month, the system also prints out final statistical summary
reports of the previous month’s QC data at the time of the Auto Clean.

1.

Access the System Options screen from the Menu screen:

a. Press

6 System Setup

and press

Enter

.

b. Press

5 Systems Options

and press

Enter

.

The System Options screen appears, as shown in Figure 5-18.

Figure 5Ć18.ā System Options Screen for an 850

Move to the Auto Clean

Time field and enter the

time.

2.

Move to the Auto Clean Time field.

3.

Type the hour and minutes in the format, HH:MM. You can enter 0 – 23 hours
and 0 – 59 minutes.

4.

Press

Done

when you finish.

5.

Define another setup function or press

Exit Menu

to return to the Ready

screen.

Scheduling Maintenance Tasks

Use this procedure to:

S define the start date for scheduling maintenance on your system
S increase or decrease the frequency of the maintenance tasks
1.

Access the Maintenance Setup screen from the Menu screen:

a. Press

6 System Setup

and press

Enter

.

6

Menu Code

5

6

Menu Code

2