Bayer HealthCare Rapidlab 800 User Manual

Page 453

Advertising
background image

System Administration

5Ć43

1.

Access the Correlation Coefficients screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

8 Correlation

and press

Enter

.

The Correlation Coefficients screen appears, as shown in Figure 5-26.

Figure 5Ć26.ā Correlation Coefficients Screen for an 850

Press Clear Entry to

delete text in a field.

2.

Type the slope and offset values you want for each parameter.

3.

Press

Done

when you finish.

4.

Define another setup function or press

Exit Menu

to return to the Ready

screen.

If you press Done before completing all the fields, the Incomplete Data Entry
message appears. Press

OK

and complete the required fields.

If you type an invalid value in a field and press Done, the Invalid Entry message
box appears. Press

OK

and type a valid entry in the field.

pH values on this screen are always expressed in pH units and not in nmols/L.

5

Menu Code

8

Procedural

Notes