Bayer HealthCare Rapidlab 800 User Manual

Page 160

Advertising
background image

2Ć62

Rapidlab 800 Operator's Manual

Figure 2Ć34.ā Calibration Data Search Log Screen for an 850

Press this key to print

or transmit a report.

NOTE:ĄĂ

The system uses two sequence numbers for two-point calibrations.

Only one number appears in the log.

5.

Perform the appropriate task at the Calibration Data Search Log.

If you want to . . .

Then . . .

view the data for a
calibration

select the report you want and press

Enter

. The

Calibration Data Search Result screen appears, as shown
in Figure 2-35.

print a calibration report

a. Select the report you want.

b. Press

Reporting Options

.

c. Press

OK.

The calibration report prints, as shown in

Figure 2-36.

transmit the results to an
LIS or a data management
system

a. Select the report you want.

b. Press

Reporting Options

.

c. Select Send Results and press

Enter

.

d. Press

OK

.