Bayer HealthCare Rapidlab 800 User Manual

Page 222

Advertising
background image

3Ć52

Rapidlab 800 Operator's Manual

Figure 3Ć33.āăRemoving a Roller Cage

Platen

Roller Cage

Roller Cage Shaft

Large Tubing

Small Tubing

Tubing Cuff

WARNING

Wear safety glasses, gloves, and a laboratory coat when handling

bleach.

6.

Clean the roller cage:

a. Clean the rollers with a lint-free tissue moistened with a 10% solution of

household bleach.

b. Clean the interior surface of the roller cage with a lint-free swab moistened

with the bleach solution.

c. Clean the roller cage shaft with a lint-free tissue moistened with the bleach

solution.

d. Rinse the rollers, the cage, and the shaft with reagent water.

e. Dry the roller cage thoroughly.

f. Ensure that the rollers turn freely.

7.

Reinstall the roller cage:

a. Replace the roller cage on the shaft.

b. Turn the roller cage on the shaft until it stops.

c. Press the roller cage down until the cage snaps into place.

8.

Connect the tubing as described in

Replacing the Pump Tubing

, page 3-39.