Replacing the coćox sample tubing – Bayer HealthCare Rapidlab 800 User Manual

Page 205

Advertising
background image

3Ć37

Maintaining the System

Replacing the COĆox Sample Tubing

Materials required:

S CO-ox sample tubing

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Replace the CO-ox sample tubing as shown in Figure 3-22:

Figure 3Ć22.āăReplacing the COĆox Sample Tubing

Sample Chamber

CO-ox Pump

Waste Tubing

Reagent Pump

Hemolyzer

FD5

Sample
Connector

Inlet

Outlet

Bubble Trap

a. Remove the anvil cap from the hemolyzer.

b. Disconnect both pieces of tubing from the anvil.

c. Remove the tubing guide on FD5 and gently pull the tubing through FD5

toward the sample connector.

d. Disconnect the CO-ox sample tubing from the sample connector.

e. Connect the flared end of the replacement tubing to the sample connector.

f. Press the tubing guide on FD5 into place.

g. Thread the replacement tubing through FD5.

h. Connect the replacement tubing to the anvil.

i. Disconnect both pieces of sample tubing from the sample chamber.

j. Connect one end of the replacement tubing to the inlet on the sample

chamber and the other end to the anvil.

k. Reinstall the anvil cap on the hemolyzer. Listen for a click, which indicates

that the anvil is in place.