Bayer HealthCare Rapidlab 800 User Manual

Page 291

Advertising
background image

4Ć9

Troubleshooting the System

System Message

Description

No Sample Device
Detected

Appears when you press Analyze, the sample door closes, and no
device is detected in the sample port. See Troubleshooting System
Messages,
page 4-120.

No Waste Bottle
Detected

Appears when the waste bottle is not installed.

__ Not Sent

Appears when you press Do Not Send at the end of analysis.

Out of Range

Appears when results for a parameter are out of the measurement
range.

Probe Detected
Obstruction

Appears when the probe hits an obstruction such as the plunger
because there is insufficient sample.

Report Results with
Cal Drift

Appears when results are reported for a sensor that exhibits
calibration drift.

Sample
Temperature
Out of Range

Appears when the measurement block temperature is not in range
at the end of measurement sequence. See Troubleshooting System
Messages,
page 4-120.

__ Sent

Appears when you press Send at the end of analysis or if results
are sent automatically.

SulfHb > 1.5%

Appears when the CO-ox module detects a concentration of
SulfHb greater than 1.5%.

tHb Slope
Calibration Due

Appears when you need to perform a tHb Slope calibration for
the CO-ox module.

Unexpected Device
Detected

Appears if the system detects an unexpected device, such as
when a sample device is inserted as the door attempts to close to
perform an automatic sequence.

Waste Full

Appears when the waste bottle can no longer accept waste.

Waste Is Almost
Full

Appears when the waste bottle is near capacity and can accept
waste from three more samples.