Analyzing vacuum tube samples – Bayer HealthCare Rapidlab 800 User Manual

Page 116

Advertising
background image

2Ć18

Rapidlab 800 Operator's Manual

Analyzing Vacuum Tube Samples

The following table lists the minimum volumes required by each system to analyze
vacuum tube samples

:

Table 2Ć5.āăMinimum Sample Volumes

System

Minimum Sample Volume

840
844, 845

0

90

µ

L

140

µ

L

850
854, 855

110

µ

L

160

µ

L

860
864, 865

125

µ

L

175

µ

L

If you want the system to analyze a different set of parameters for the sample,
press Change Parameters to select another panel. Refer to Selecting Parameter
Panels
, page 2-29, for more information.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Prepare the sample and insert an aspiration adapter into the sample port.

2.

Insert the adapter into the vacuum tube, as shown in Figure 2-12.

Figure 2Ć12.ā Inserting an Aspiration Adapter with a Vacuum Tube Sample

Sample Port

Vacuum Tube

Aspiration Adapter