Defining automatic transmission of results – Bayer HealthCare Rapidlab 800 User Manual

Page 448

Advertising
background image

800 Series Operator's Manual

5Ć38

Defining Automatic Transmission of Results

Use this procedure to select automatic transmission of patient sample results
directly to an LIS, HIS, or data management system. You can use this option only
if your 800 system is connected to an LIS or data management system. Refer to
Selecting Automatic Transmission of QC Results, page 5-55, for the procedure to
define auto send for QC results.

Table 5Ć12.āăAuto Send Patient Sample Options

Option

Description

On

The system automatically sends patient sample results to an LIS or data
management system. The default setting is On.

Off

Results are not automatically transmitted. The operator must press

Send

to

transmit results or press

Do Not Send

to prevent transmission of results.

1.

Access the Systems Options screen from the Menu screen:

a. Select

6 System Setup

and press

Enter

.

b. Select

5 Systems Options

and press

Enter

.

The System Options screen appears, as shown in Figure 5-23.

Figure 5Ć23.ā System Options Screen for an 850

Move to the Auto Send

frame and select

Patient.

2.

Move to the Auto Send frame.

3.

Select

Patient

and press

Enter

.

A dark box indicates the option is on.

4.

Press

Done

.

5.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

6

Menu Code

5

Advertising