800 Series Operator's Manual

Removing the Gas Tanks

Compressed gas tanks require cautious handling. To prevent

damage and possible personal injury, refer to Replacing the Gas Tanks in Section
3, for more detailed precautions.

1.

Remove the gas tanks, as shown in Table H-4:

Table HĆ4.āăRemoving the Gas Tank and Regulator

Needle Valve

Adjustment Knob

Gas Tank Seal

Regulator Nipple

Yoke

Screw

Gas Line

Yoke

Dowel Pins

Valve Stem

Valve Outlet

Dowel Pin Holes

Gas Tank

a. Using a valve wrench, close each gas tank by turning the valve stem

clockwise.

b. Disconnect the gas regulators from the gas tanks by unscrewing the yoke

screws.

c. Verify that the gas tank seals are in place on the regulators.

d. Remove the gas tanks to a well ventilated, open area.

e. Position each valve outlet so that it is facing down, away from loose

objects.

CAUTION:ĄĂ

Avoid contact with the gas stream. Gas under pressure can cause

bodily injury and property damage.

f. Using the wrench, release the contents of the gas tanks by turning the valve

stems counterclockwise.

WARNING

!