EĆ20

800 Series Operator's Manual

Table E-16 through Table E-22 summarize the results of the 850 system whole
blood, expired gas, and electrolyte recovery and precision testing.

Table EĆ16.āă850 Recovery and Precision TestingĊpH

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

24

0.002

7.194

7.193

100.0

Capillary

24

0.003

7.194

7.194

100.0

Microcapillary

24

0.003

7.170

7.175

99.9

Microsyringe

24

0.002

7.229

7.225

100.1

pH/Lytes

18

0.003

7.220

7.218

100.0

2

Syringe

24

0.001

7.399

7.405

99.9

Capillary

24

0.002

7.397

7.393

100.1

Microcapillary

24

0.003

7.402

7.396

100.1

Microsyringe

24

0.002

7.383

7.397

99.8

pH/Lytes

18

0.004

7.357

7.356

100.0

3

Syringe

24

0.004

7.493

7.493

100.0

Capillary

24

0.003

7.502

7.505

100.0

Microcapillary

24

0.006

7.449

7.459

99.9

Microsyringe

24

0.003

7.557

7.560

100.0

pH/Lytes

18

0.005

7.673

7.676

100.0

* WRSD = within-run standard deviation