Removing the tubing tab from the 800 base model, Updating the system id – Bayer HealthCare Rapidlab 800 User Manual

Page 611

Advertising
background image

Installation

HĆ19

NOTE:

Ensure that the waste tubing is positioned through the hole on the CO-ox

module fluid detector.

5.

Connect the waste tubing to the 800 base model manifold.

6.

Connect the sample tubing to the sample connector.

Installing the Printer Cover with Correct Model Number

This procedure refers to right and left as you look at the front of system.

1.

Remove the printer paper.

2.

Remove the old printer cover:

a. Lift the cover and slide it to the right until the cover flange clears the pivot

pin.

b. When the left side of the cover is free of the pivot pin, pull the cover to the

left to remove it from the printer.

3.

Install the new printer cover:

a. Slide the left tab over the pin on the spindle arm bracket while pushing the

cover all the way to the left.

b. Bend the tab on the cover or pivot it to allow the tab to slide over the pivot

pin.

Removing the Tubing Tab from the 800 Base Model

1.

Locate the groove on the left side of the base model front cover.

2.

Using pliers, grasp the tubing tab inside the grooved area. Move and flex the
tubing tab to weaken the joint until the tab breaks free from the cover.

3.

Close the system:

a. Tighten both locking screws on the base unit.

b. Tighten the one locking screw on the CO-ox module.

Updating the System ID

This procedure is used by Bayer Diagnostics Service Representatives to change the
system information that the 800 system uses to present information on the screen
and in printed reports.

NOTE:

A service password is required to access this setup function.

1.

Reconnect all cables to the ports as shown in Figure H-5.