Troubleshooting the System

5.

Perform the appropriate action.

If you want to . . .

Then . . .

test another reagent

repeat steps 2 through 4.

return to the Menu
screen

press

Exit Test

.

6.

Press

Exit Menu

.

Pump Flow Rate Test

Use this procedure to test the flow rate of the reagent pump and to calibrate the
reagent pump, if necessary.

1.

Access the Pump Flow Rate Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

2 Pump Flow Rate

and press

Enter

.

The Pump Flow Rate Test screen appears as shown in Figure 4-3.

Figure 4Ć3.ā Pump Flow Rate Test Screen

The flow rate units are mL/sec.

Check here for results.

2.

Press

Start Test

.

2

1

Menu Code

3