Bayer HealthCare Rapidlab 800 User Manual

Page 354

Advertising
background image

4Ć68

800 Series Operator's Manual

Figure 4Ć9.ā Sample Entry Test Screen

Perform the automatic

test to check the

sample door and

sample probe positions.

Check here for results.

2.

Test the positions of the sample door and the sample probe:

a. Select

Automatic

and press

Enter

.

b. Press

Start Test

.

c. Check the screen for the message, Sample entry module acceptable, to

verify that the sample door and sample probe function correctly.

d. Press

Exit Test

.

3.

Test the ability of the system to recognize the sample device:

a. Select

Manual

and press

Enter

.

b. Insert a sample device, with the plunger pulled back, into the sample port

when prompted.

c. Press

Start Test

.

d. Remove the sample device when prompted.

e. Check the screen to verify that the system correctly identifies the size and

the type of the sample device.

f. Press

Exit Test

.

4.

Test the ability of the sample probe to detect obstructions:

a. Select

Manual

and press

Enter

.

b. Push in completely the plunger of a syringe and insert the syringe into the

sample port when prompted.

c. Press

Start Test

.

d. Remove the syringe when prompted.

e. Check the screen to verify that the sample probe detects an obstruction and

identifies the size of the sample device.