Troubleshooting the System

3.

Inspect the sample connector for obstructions:

a. Disconnect the tubing from the sample connector.

b. Remove the sample connector.

c. Push a clot-removal line through the sample connector to remove

obstructions.

NOTE:ĄĂ

If the obstruction is in the preheater, see Removing Obstructions from

the Measurement Module for instructions.

d. Reinstall the sample connector.

e. Reconnect the tubing.

4.

Inspect the hemolyzer for obstructions:

a. Remove the anvil cap on the hemolyzer by turning it a quarter turn

counterclockwise.

b. Disconnect the sample tubing from the anvil.

c. Pull the anvil and anvil spring away from the mounting pin on the

hemolyzer.

d. Push a clot-removal line through the tubing connectors on the anvil.

e. Clean the exterior surface.

f. Reassemble the hemolyzer.